Furnaces
Thermolyne* Temperature Furnace Control Information

Temperature Control: Precision and Accuracy


  • The temperature control system in Barnstead Thermolyne furnaces consists of four major components: controller, power-switching device, heating elements, and a thermocouple sensor (Figure 1).

  • All of the Barnstead Thermolyne furnace control systems are closed-loop for accuracy and safety.

  • In a closed-loop system, the temperature sensed by the thermocouple gives a feedback signal to the controller that in turn regulates power to the heating elements by means of the switching device.

  • The selection of components within the control loop for a particular furnace model is largely based on the desired operating temperature range and the magnitude of overall temperature error that is acceptable within that range.

  • The overall error is defined as the difference between the set value and the actual value of the furnace chamber or load.

  • The overall error is broken down into the precision of the temperature reading and its accuracy.

  • Temperature precision is largely determined by the controller.

  • Our economy solid tube furnaces use analog controllers and all other Barnstead Thermolyne furnaces utilize microprocessor (digital) three-term controllers (Proportional + Integral + Derivative or PID).

  • PID control is very precise in that small fluctuations in chamber temperature are quickly adjusted to maintain the set temperature.

  • In addition, the PID type controller reaches the desired chamber temperature and reduces the chance of overshooting the value.

  • This is particularly important to consider if the furnace workload is sensitive to temperatures slightly above the setpoint.

  • Temperature control in furnaces is the most precise above 260°C.

  • Other factors to consider with regard to accuracy include the size and density of the material being heated and the thermal lag of the material relative to the chamber temperature.

  • Large, dense materials will obviously require more time to reach the set temperature.

  • Over-Temperature Protection (OTP)


  • Digital controllers are designed with two types of OTP.

  • A "full-scale high" alarm will be in effect when the furnace chamber exceeds its maximum allowable temperature.

  • The controller will cut power to the heating elements until the chamber cools to a value below this limit.

  • This value is defined by the heating capacity of the elements, the thermocouple, and the insulation and cannot be modified by the operator.

  • The second type of OTP, the "deviation high" alarm is designed to protect the furnace workload and can be changed by the operator.

  • We recommend setting this value to 20°C above the working temperature of the load.

  • Sensor Break Protection


  • All Barnstead Thermolyne controllers provide for a safe response should there be a loss of input signal because of a broken thermocouple or an open circuit.

  • Power to the heating elements is terminated and a message flashes on the display.

  • Tuning


  • The digital controllers can be automatically tuned to improve product or process quality.

  • Self-tuning is a built-in software package for automatically determining the PID and other control parameter values.

  • For maximum performance, this feature should be used the first time the furnace is used and each time the setpoint or the type of load is changed.

  • Programming


  • Many processes require the temperature to vary according to a precise profile with time.

  • The digital controllers allow a programmed function which is stored in non-volatile memory.

  • Programs consist of functions and associated settings called segments.

  • These segments can be assembled into a program in a fixed format, such as a succession of ramp/dwell pairs (Figure 2).

  • Segments include: Ramp: The setpoint temperature increases or decreases at a linear ramping rate until a specified target level is reached.

  • Step: A step is a specialized type of ramp segment where the setpoint jumps to the specified target level.

  • It is used when the most rapid change in measured value, limited by the heating elements, is desired.

  • Dwell: The length of time required at a target temperature.

  • Communications


  • Bi-directional digital communications via an RS-232 port is an option with select digital controllers.

  • Eurotherm Controls has available a free demonstration program called the Communications Sampler.

  • This program enables Windows® to communicate with a single controller.

  • The controller's parameters can be displayed on the screen and the value of the parameters can be changed with the computer keyboard.

  • Two off-the-shelf software packages, Eurotherm’s iTools and SpecView allow the user to supervise groups of controllers, as well as retrieve data for logging and trend plots.

  • Holdback


  • If the measured value lags behind the setpoint by a preset amount during ramp or dwell, the holdback feature can be used to freeze the program at its current state.

  • The program will resume when the error comes within the holdback value.

  • Gain Scheduling


  • Gain scheduling is the automatic transfer of control between two sets of PID values.

  • Select digital controllers do this at a preset measured temperature.

  • Gain scheduling is used for difficult to control processes which show large changes in their response time or sensitivity at high or low temperatures.









  • All trademarks are not owned by Alpha Multiservices, Inc, they are property of their respective owners. Prices and availability of products listed are subject to change or may be discontinued without notice. Product descriptions, pictures and availability are provided with intent of accuracy; however, Alpha Multiservices, Inc. is not liable for errors, omissions or misrepresentations. If you are experiencing technical difficulties, please contact the site producer (webmaster).

    * Tel: 936.321.4582
    * Toll: 877.530.3232
    * Fax: 936.273.4584
    * support @ alphamultiservices.com
    * webmaster @ alphamultiservices.com
    Aquametry Apparatus
    Balances/Scales
    Centrifuges
    Cryogenics
    Deionization
    Dispensers & Dilutors
    Distillation
    Dri Baths
    Electrophoresis
    Environmental Chambers
    Filtration
    Fluorometers
    Furnaces
    Heating Blocks
    Heating Mantles
    Histology
    Hot Plates
    Incubators
    Laboratory Accessories
    Laboratory Apparatus
    Melting Point
    Moisture Analyzers
    Ovens
    Pyrometers
    Recorders
    Refrigerators/Freezers
    Reverse Osmosis
    Shakers
    Spectrophotometers
    Steaming Baths
    Sterilizers
    Stirrers
    Stirring Hot Plates
    Thermal Cyclers
    Thermometers
    Ultrasonic Cleaners
    Vacuum Pumps
    Viscosimeters
    Vortex Mixers
    Water Baths

    Copyright 2009 Alpha Multiservices, Inc.- All Rights Reserved